PTA球囊擴張導管中2-氯乙醇殘留量的評價方法

Evaluation method of residual ethylene chlorohydrin in PTA balloon dilatation catheter

作者: 付步芳  程東升  王春仁 
單位:<span style=";font-family:宋體">中國食品藥品檢定研究院 (北京 102629)</span><p><span style=";font-family:宋體">煙臺毓璜頂醫(yī)院 (山東煙臺 264000)</span></p>
關鍵詞: PTA球囊擴張導管;2-氯乙醇;殘留量;氣相色譜-質譜 
分類號:<span style=";font-family:宋體">R318.08</span>
出版年·卷·期(頁碼):2018·37·1(62-65)
摘要:

Objective To develop a gas chromatography-mass spectrometric(GC-MS) method for the determination and evaluation of ethylene chlorohydrin (ECH)  residue in PTA balloon dilatation catheter. Methods The ECH residue were extracted in acetone by ultrasonic bathing. Chromatographic column: ZB-624, inlet temperature 180, split ratio 1:1, high purity He as carrier gas, column flow rate 1.5mL/min. Temperature programmed condition: 50 (1min), 5/min to 100; Mass spectrometric conditions were as follow: ion source temperature 230℃,the interface temperature between GC and MS was 250, electrical energy 70 eV. Qualitative analysis was carried out by retention time and mass spectrometric scan , m/z between 35500, and quantified by select ion monitoring method (SIM), m/z=43. The ECH was determined by external standard method under the SIM mode by GC-MS. Results The number of theoretical plates were more than 56753. A good linearity was obtained at the concentrations of 1.44~22.96 μg/mL with a correlation of 0.999 and the average recovery was 99.1% (n=5, RSD<5%). The minimum limit of ECH was 0.41μg/mL. No ECH was found in the sample. Conclusions The method of GC-MS is sensitive, rapid and accurate for determination of ECH residue in PTA balloon dilatation catheter products. It can be used for quality control of PTA balloon dilatation catheter and for peers reference.

參考文獻:

服務與反饋:
文章下載】【加入收藏
提示:您還未登錄,請登錄!點此登錄
 
友情鏈接  
地址:北京安定門外安貞醫(yī)院內北京生物醫(yī)學工程編輯部
電話:010-64456508  傳真:010-64456661
電子郵箱:[email protected]